Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial

Myers MG, Godwin M, Dawes M, Kiss A, Tobe SW, Grant FC, Kaczorowski J.
BMJ 2011;342:d286


OBJECTIVE: To compare the quality and accuracy of manual office blood pressure and automated office blood pressure using the awake ambulatory blood pressure as a gold standard.
DESIGN: Multi-site cluster randomised controlled trial.
SETTING: Primary care practices in five cities in eastern Canada.
PARTICIPANTS: 555 patients with systolic hypertension and no serious comorbidities under the care of 88 primary care physicians in 67 practices in the community.
INTERVENTIONS: Practices were randomly allocated to either ongoing use of manual office blood pressure (control group) or automated office blood pressure (intervention group) using the BpTRU device. The last routine manual office blood pressure (mm Hg) was obtained from each patient's medical record before enrollment. Office blood pressure readings were compared before and after enrollment in the intervention and control groups; all readings were also compared with the awake ambulatory blood pressure.
MAIN OUTCOME MEASURE: Difference in systolic blood pressure between awake ambulatory blood pressure minus automated office blood pressure and awake ambulatory blood pressure minus manual office blood pressure.
RESULTS: Cluster randomisation allocated 31 practices (252 patients) to manual office blood pressure and 36 practices (303 patients) to automated office blood pressure measurement. The most recent routine manual office blood pressure (149.5 (SD 10.8)/81.4 (8.3)) was higher than automated office blood pressure (135.6 (17.3)/77.7 (10.9)) (P < 0.001). In the control group, routine manual office blood pressure before enrollment (149.9 (10.7)/81.8 (8.5)) was reduced to 141.4 (14.6)/80.2 (9.5) after enrollment (P < 0.001/P = 0.01), but the reduction in the intervention group from manual office to automated office blood pressure was significantly greater (P < 0.001/P = 0.02). On the first study visit after enrollment, the estimated mean difference for the intervention group between the awake ambulatory systolic/diastolic blood pressure and automated office blood pressure (-2.3 (95% confidence interval -0.31 to -4.3)/-3.3 (-2.7 to -4.4)) was less (P = 0.006/P = 0.26) than the difference in the control group between the awake ambulatory blood pressure and the manual office blood pressure (-6.5 (-4.3 to -8.6)/-4.3 (-2.9 to -5.8)). Systolic/diastolic automated office blood pressure showed a stronger (P < 0.001) within group correlation (r = 0.34/r = 0.56) with awake ambulatory blood pressure after enrollment compared with manual office blood pressure versus awake ambulatory blood pressure before enrollment (r = 0.10/r = 0.40); the mean difference in r was 0.24 (0.12 to 0.36)/0.16 (0.07 to 0.25)). The between group correlation comparing diastolic automated office blood pressure and awake ambulatory blood pressure (r = 0.56) was stronger (P < 0.001) than that for manual office blood pressure versus awake ambulatory blood pressure (r = 0.30); the mean difference in r was 0.26 (0.09 to 0.41). Digit preference with readings ending in zero was substantially reduced by use of automated office blood pressure.

 

BMJ 2011;342:d286

 


Area Soci

Eventi

38° Congresso Nazionale


38° Congresso Nazionale

Bologna, 1-3 dicembre 2024

[continua a leggere]

Congresso Regionale SISA Sezione Campania

Napoli, 13 Dicembre 2023

[continua a leggere]

Congresso Regionale SISA Sezione Siculo-Calabra

Catania, 3-4 Dicembre 2023

[continua a leggere]

SISA LIPID ACADEMY - Corso avanzato di lipidologia clinica

Modena, 22-23 Giugno 2023

[continua a leggere]

Giornale Italiano Arteriosclerosi

Rivista in lingua italiana
riservata ai Soci SISA
Ultimo numero:
Anno 14 • N.4/2023

[continua a leggere]

HoFH today

Rivista Italiana della
Ipercolesterolemia
Familiare Omozigote
Anno 5 • N.1/2023

[continua a leggere]

Rivista NMCD

Nutrition, Metabolism and Cardiovascular Diseases

Istruzioni per l'accesso online

IF 2018: 3.340


Diateca

EAS Lipid Clinic Webinar - Le linee guida per le dislipidemie: presente e futuro
[continua a leggere]
EAS Advanced Course in Rare Lipid Disease
[continua a leggere]

Newsletter

Per essere informati sulle novità di S.I.S.A. iscrivetevi alla nostra newsletter inserendo
il vostro indirizzo di posta elettronica

Progetto LIPIGEN

LIPIGEN
Nuovo sito dedicato al Progetto LIPIGEN

 

Progetto LIPIGEN - Vecchio portale
E' necessario essere loggati come utente
Lipigen per poter accedere alla pagina

PROject Statin Intolerance SISA

 

PROSISA – PROject Statin Intolerance SISA
E' necessario essere loggati come utente
PROSISA per poter accedere alla pagina

GILA - Lipoprotein Aferesi

 

Gruppo Interdisciplinare Lipoprotein Aferesi
(Accesso Gruppo GILA-Lipoprotein Aferesi)

E' necessario essere loggati come utente del Gruppo GILA per poter accedere

 

Gruppo Interdisciplinare Lipoprotein Aferesi
(Documentazione ad accesso libero)

Pagina informativa per medici e pazienti